Skip to content

Public FDA record

Z-1352-2019: Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 18300

Recall record Z-1352-2019 for Moller Vital. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1352-2019
Company in FDA record
Moller Vital
FDA record creation date
May 16, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.

Source facts

Action
On February 4, 2019, the firm notified customers of the recall via Urgent Field Safety Notice letter. The letter alerted customers to the recalled product. The firm stated that all new 2-Point Sling Bars are now equipped with metal alloy hooks instead of the previously used composite hooks. As part of this recall, the firm will replace the recalled devices with the new design. Distributors were asked to do the following: 1) Read the instructions, and complete and return the customer response form 2) Notify affected customers of the recall 3) Review the list of sales records from Etac Supply Gjvik and supply additional information from your records. 4) Identify and contact end-customers/users and decide the requirement for ordering new units 5) Order new 2-Point Sling Bars from Etac Supply Gjvik 6) Ship replacement units to end customers accompanied by replacement instructions (see user manual) and the Consignee Form. The End-User/Customer must perform the replacement and fill the replacement records (green fields) in the consignee form and return a signed copy to the distributor. 7) Receive the signed Consignee Form from end-users/customers. Perform weekly reporting to Etac Supply Gjvik in the same spreadsheet that contains the sales records. The signed Consignee Forms must be returned to Etac Supply Gjvik as a complete package upon finalizing all replacements.
Initiated
2019-02-04
Root cause
Use error
Distribution
Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
Recall number
Z-1352-2019
Recall status
Open, Classified
CDRH record ID
170521
FDA record created
2019-05-16
Recall event number
81977

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗