Open original FDA record ↗Recall record
- FDA identifier
- Z-1352-2020
- Company in FDA record
- Breas Medical, Inc.
- FDA record creation date
- Feb 24, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
Source facts
- Action
- On January 28, 2020, the firm began distributing Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue and associated risk of hypercarbia caused by excessive CO2 build-up (rebreathing). Customers were instructed to do the following: " Identify users/patients who use Protective Covers (p/n 006344) with a Vivo 50 or Vivo 65 ventilators. The model of ventilator can be determined from the device label on the bottom cover. Vivo 50 units have part no 215000 and Vivo 65 units have part no 226500. " Check if the Protective Cover is used with an exhalation valve circuit or dual-limb circuit. If this is the case, the users should be advised to take note of Rebreathing alarms and Exhalation Valve Control Error alarms. If any of these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This may be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. " Please notify Breas Medical of your preferred solution from the options below. " Please fill out and return the attached Reply Form at your earliest convenience and latest 29 February 2020. " Please forward this notice to anyone to whom you may have further distributed the product (additional consignees). Breas offers two options/solutions, which can be combined. 1. Advise users of the Protective Covers to be observant of the Rebreathing alarms and Exhalation Valve Control Error alarms and inform them that if these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This can be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. With this precaution, the Protective Covers can continue to be used. 2. If a user prefers, Breas offers to replace Protective Covers (p/n 006344) with new Protective Covers free-of-charge. We can provide you up to the number of the Protective Covers purchased according to our reco
- Initiated
- 2020-01-28
- Root cause
- Device Design
- Terminated
- 2020-09-15
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.
- Recall number
- Z-1352-2020
- Recall status
- Terminated
- CDRH record ID
- 179485
- FDA record created
- 2020-02-24
- Recall event number
- 84828