Z-1352-2020: Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective c

Recall record Z-1352-2020 for Breas Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1352-2020
Company in FDA record
Breas Medical, Inc.
FDA record creation date
Feb 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.

Source facts

Action
On January 28, 2020, the firm began distributing Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue and associated risk of hypercarbia caused by excessive CO2 build-up (rebreathing). Customers were instructed to do the following: " Identify users/patients who use Protective Covers (p/n 006344) with a Vivo 50 or Vivo 65 ventilators. The model of ventilator can be determined from the device label on the bottom cover. Vivo 50 units have part no 215000 and Vivo 65 units have part no 226500. " Check if the Protective Cover is used with an exhalation valve circuit or dual-limb circuit. If this is the case, the users should be advised to take note of Rebreathing alarms and Exhalation Valve Control Error alarms. If any of these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This may be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. " Please notify Breas Medical of your preferred solution from the options below. " Please fill out and return the attached Reply Form at your earliest convenience and latest 29 February 2020. " Please forward this notice to anyone to whom you may have further distributed the product (additional consignees). Breas offers two options/solutions, which can be combined. 1. Advise users of the Protective Covers to be observant of the Rebreathing alarms and Exhalation Valve Control Error alarms and inform them that if these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This can be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. With this precaution, the Protective Covers can continue to be used. 2. If a user prefers, Breas offers to replace Protective Covers (p/n 006344) with new Protective Covers free-of-charge. We can provide you up to the number of the Protective Covers purchased according to our reco
Initiated
2020-01-28
Root cause
Device Design
Terminated
2020-09-15
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.
Recall number
Z-1352-2020
Recall status
Terminated
CDRH record ID
179485
FDA record created
2020-02-24
Recall event number
84828

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗