- FDA identifier
- Z-1353-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Feb 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Issue with software algorithm which may lead to overpressure events.
Source facts
- Action
- Olympus notified consignees on about 01/16/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to cease usage of the affected product immediately. They are advised to use available alternatives including an existing UHI-4 if available or switching to an insufflation device from another manufacturer. If no alternative is available on site, contact Olympus to arrange a loaner or purchase a replacement. Additionally, consignees were instructed to examine inventory and quarantine any identified devices, immediately cease usage of any UHI, UHI-2, and/or UHI-3 in inventory, call Olympus Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return your device. Consignees were also requested to acknowledge receipt of this letter through their recall web portal as follows: a. Go to https://olympusamerica.com/recall, b. Enter the recall number "0489 and c. Complete the form as instructed. Please indicate in your response if you no longer have the device(s). If any consignees have further distributed this UHI product, they were instructed to identify affected customers and forward this notice to them immediately. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by email at Cynthia.Ow@Olympus.com.
- Initiated
- 2026-01-16
- Root cause
- Software design
- Distribution
- Nationwide distribution
- Recall number
- Z-1353-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218005
- FDA classified
- 2026-02-18
- Recall event number
- 98243