Open original FDA record ↗Recall record
- FDA identifier
- Z-1354-2018
- Company in FDA record
- C.R. Bard, Inc.
- FDA record creation date
- Apr 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
Source facts
- Action
- The firm initiate their recall by letter on 02/27/2018 by letter. The recall letter stated: "Action required: " Immediately examine your inventory and quarantine product subject to this recall. Please refer to the sample product labeling on Attachment 1 to help you locate the product. Do not use or further distribute any affected product. " Please complete and return the accompanying Recall & Effectiveness Check Form attached to this letter regardless of whether or not you have any of the affected product. Additional instructions for product return are contained on the form. Upon receipt of the form indicating there is product to return, the BMD Recall Coordinator will issue you a return authorization number for return of the affected product. " If you have further distributed any units from the affected lots, please identify your customers and notify them at once of this product recall by forwarding a copy of this letter. Your notification should include a copy of this letter and the accompanying enclosures. This recall should be carried out to the user level."
- Initiated
- 2018-02-27
- Root cause
- Finished device change control
- Terminated
- 2019-07-11
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1354-2018
- Recall status
- Terminated
- CDRH record ID
- 162513
- FDA record created
- 2018-04-11
- Recall event number
- 79507