905 nm laser diodes may be below rated power output.
Source facts
Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
.Healthcare practitioners are instructed to courier their TLC-2000 to Theralase for analysis. Their 905 nm laser diodes will be analyzed and recalibrated, as required, to meet rated power output.
Initiated
2019-09-26
Root cause
Device Design
Terminated
2020-06-04
Distribution
USA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Recall number
Z-1354-2020
Recall status
Terminated
CDRH record ID
179528
FDA classified
2020-02-13
Recall event number
84843