Open original FDA record ↗Recall record
- FDA identifier
- Z-1354-2024
- Company in FDA record
- SleepNet Corporation
- FDA classification date
- Apr 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Update to contraindications and warning language due to CPAP masks containing magnets.
Source facts
- Action
- The firm emailed consignees a notification titled URGENT: MEDICAL DEVICE RECALL, dated 3/1/24. Consignees were instructed to stop using recalled masks if the user or anyone in close physical contact with the mask has an active medical implant or metallic implant that will interact with magnets and contact their mask provider for a replacement mask that does not contain magnets. Consignees that do not have an active medical implant or metallic implant that will interact with magnets do not have to cease use. If a consignee is unsure whether or not to cease use, they should consult with their physician or manufacturer of the implant. Consignees are to continue using masks according to the updated instructions and labeling if patients or people in close proximity to them do not have implanted metallic medical devices or metallic objects in the body. Consignees with any questions are to contact Himabindu Koppusetty via email at hkoppusetty@sleepnetcorp.com or by phone at 1-800-742-3646x6611 from Monday through from, 8:00 AM to 4:00 PM.
- Initiated
- 2024-03-01
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of AUSTRIA, BAHRAIN, BRAZIL, COLOMBIA, COSTA RICA, ECUADOR, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, INDIA, IRAQ, IRELAND, ISRAEL, ISREAL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SULTANATE OF OMAN, SWEDEN, SWITZERLAND, THE NETHERLANDS, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VIETNAM, YEMEN.
- Recall number
- Z-1354-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206508
- FDA classified
- 2024-04-01
- Recall event number
- 94169