Open original FDA record ↗Recall record
- FDA identifier
- Z-1355-2020
- Company in FDA record
- Cook Inc.
- FDA classification date
- Jan 31, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.
Source facts
- Action
- On January 14, 2020, the firm notified affected customers via Urgent Medical Device Correction letters/emails. Customers were advised to pay especial attention to several aspects of the Instructions for Use (IFU) which were of key importance when implanting a device. Customers were also provided with an updated Planning and Sizing Worksheet. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
- Initiated
- 2020-01-16
- Root cause
- Labeling design
- Terminated
- 2020-11-12
- Distribution
- Foreign distribution only. Distributed worldwide.
- Recall number
- Z-1355-2020
- Recall status
- Terminated
- CDRH record ID
- 179350
- FDA classified
- 2020-01-31
- Recall event number
- 84778