Z-1355-2024: Mojo 2 Full Face AAV Non Vented Mask, All Sizes

Recall record Z-1355-2024 for SleepNet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1355-2024
Company in FDA record
SleepNet Corporation
FDA classification date
Apr 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Update to contraindications and warning language due to CPAP masks containing magnets.

Source facts

Action
The firm emailed consignees a notification titled URGENT: MEDICAL DEVICE RECALL, dated 3/1/24. Consignees were instructed to stop using recalled masks if the user or anyone in close physical contact with the mask has an active medical implant or metallic implant that will interact with magnets and contact their mask provider for a replacement mask that does not contain magnets. Consignees that do not have an active medical implant or metallic implant that will interact with magnets do not have to cease use. If a consignee is unsure whether or not to cease use, they should consult with their physician or manufacturer of the implant. Consignees are to continue using masks according to the updated instructions and labeling if patients or people in close proximity to them do not have implanted metallic medical devices or metallic objects in the body. Consignees with any questions are to contact Himabindu Koppusetty via email at hkoppusetty@sleepnetcorp.com or by phone at 1-800-742-3646x6611 from Monday through from, 8:00 AM to 4:00 PM.
Initiated
2024-03-01
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of AUSTRIA, BAHRAIN, BRAZIL, COLOMBIA, COSTA RICA, ECUADOR, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, INDIA, IRAQ, IRELAND, ISRAEL, ISREAL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SULTANATE OF OMAN, SWEDEN, SWITZERLAND, THE NETHERLANDS, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VIETNAM, YEMEN.
Recall number
Z-1355-2024
Recall status
Open, Classified
CDRH record ID
206509
FDA classified
2024-04-01
Recall event number
94169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗