Z-1355-2026: da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Recall record Z-1355-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1355-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Feb 12, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Source facts

Action
On 12/30/2025, correction notices were mailed and emailed to customers who were informed the following: If single-viewer failure occurs during a procedure, when clinically safe to do so, restart the system to restore 3D visualization in the viewer. If dual-viewer failure occurs during a procedure, be aware that instrument motion will not be automatically halted by the system. Instrument motion can be halted by removing the surgeon s head from the viewer. The surgeon maintains the ability to control the instruments while their head is in the viewer. To ensure instruments do not cause unintended tissue interaction, the bedside user should maneuver installed instruments into a safe space in the surgical field, using an external monitor for visualization. Once the instruments are safely positioned, restart the system to restore visualization. At any point during the sequence, the surgeon may press E Stop to put the system in a safe state (arms locked, no instrument motion). If any viewer display issue is detected, the user should avoid starting any additional procedures with the affected system until firm has repaired the system. Forward this notification to all personnel at your site who perform and support these procedures. Complete and return the acknowledgement form via email to Recalls@intusurg.com Retain a copy of this letter and the acknowledgement form for your files. Firm's Field Engineers will contact affected customers to schedule a site visit to upgrade the Viewer Display firmware on affected systems. If you need further information or support concerning this Urgent Medical Device Correction, please contact your Clinical Sales Representative or contact the firm's Customer Service at: North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com.
Initiated
2025-12-30
Root cause
Process change control
Distribution
US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.
Recall number
Z-1355-2026
Recall status
Open, Classified
CDRH record ID
217888
FDA classified
2026-02-12
Recall event number
98205

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗