Z-1356-2023: ProxiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 706100 and 706110

Recall record Z-1356-2023 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1356-2023
Company in FDA record
Philips North America Llc
FDA classification date
Apr 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

Source facts

Action
An IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the CombiDiagnost R90 system. The system may experience a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board, with a potential for smoke and/or a popping sound coming from the Main cabinet. If a user sees or smells smoke or hears a popping noise coming from the M-cabinet, immediately remove power to the room (main circuit breaker) and remove the patient. Stop using the system for further examinations. An IMPORTANT PRODUCT NOTICE dated 10/21/21 was sent to customers with the ProxiDiagnost N90. Customers may encounter one or both issues: 1: Potential component damage due to short circuit, leading to smoke, or a popping noise coming from the M-cabinet. 2: Potential table stop due to a broken tabletop cable. An IMPORTANT PRODUCT NOTICE dated 11/15/22 was sent to customers with the ProxiDiagnost N90 1.1 for the same issues as the ProxiDiagnost N90 letter dated 10/21/21. An URGENT Medical Device Correction Notification Update for the CombiDiagnost R90 and ProxiDiagnost N90 dated February 2023 was sent to customers. For component damage due to short circuit, if a user observes fire in the M-cabinet, sees or smells smoke, or hears a popping noise coming from the M-cabinet: Contact Customer Care Solutions Center (1-800-722-9377). Reference FCO70900054 (CombiDiagnost R90), FCO70600105 (ProxiDiagnost N90 1.0), or FCO70600108 (ProxiDiagnost N90 1.1) to arrange for a Philips Field Service Engineer to visit your site. For ProxiDiagnost N90 table stop due to a broken tabletop cable, if the table is not tilting or braking as expected: Contact Customer Care Solutions Center (1-800-722-9377). Reference FCO70600105 (ProxiDiagnost N90 and N90 1.0) or FCO70600108 (ProxiDiagnost N90 1.1) to arrange for a Philips Field Service Engineer to visit your site to inspect and fix the cable. UPDATED 7/27/2023: Scope of affected products was increased. Additional consignees were notifie
Initiated
2021-10-21
Root cause
Device Design
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1356-2023
Recall status
Open, Classified
CDRH record ID
199060
FDA classified
2023-04-10
Recall event number
91853

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗