Open original FDA record ↗Recall record
- FDA identifier
- Z-1357-2015
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA classification date
- Mar 31, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This lot of the RECLAIM Assembled Implant Inserter Adaptor Instrument may fracture during surgery and there is the potential for plastic pieces to be left in the patient.
Source facts
- Action
- The firm issued a voluntary device removal on 02/23/2015 for lot no. SO2018074 of the RECLAIM Assembled Implant Inserter Adaptor. The recall notice informs US Distributors and Medical Professionals about the issue and any actions to take. US Distributors will return affected instruments and notify one Hospital/User Facility with the Hospital Notice and Reconciliation Forms. Reconciliation forms are required for the end-user Hospital/Medical Facility Consignee. Customers may contact Mindy K. Tinsley Director Wordwide Communications Tel: (574) 372-7136 Fax: (574) 372-7717 E-mail: mtinsley@its.jnj.com
- Initiated
- 2015-02-23
- Root cause
- Under Investigation by firm
- Terminated
- 2015-09-28
- Distribution
- Worldwide Distribution: US (nationwide) in CA; Ireland and Israel.
- Recall number
- Z-1357-2015
- Recall status
- Terminated
- CDRH record ID
- 134495
- FDA classified
- 2015-03-31
- Recall event number
- 70692