Z-1357-2023: Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or R

Recall record Z-1357-2023 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1357-2023
Company in FDA record
Covidien
FDA classification date
Apr 19, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

Source facts

Action
On 3/24/23, recall notices were mailed to risk managers and directors of respiratory care. Tracheostomy Tubes that do not securely attach or stay attached to caps or accessories should not be used. 1) All affected unused product should be quarantined and returned to the recalling firm. 2) Complete and then return the customer confirmation form to rs.gmbfcamitg@medtronic.com 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. 4) If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. Customers with questions can contact Quality Assurance at 800-255-6774, option 4.
Initiated
2023-03-24
Root cause
Component change control
Distribution
Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State Of Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Malta, Martinique, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic Of Tanzania, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, Viet Nam.
Recall number
Z-1357-2023
Recall status
Open, Classified
CDRH record ID
199257
FDA classified
2023-04-19
Recall event number
91943

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗