Z-1358-2024: Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Recall record Z-1358-2024 for Bio-Rad Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1358-2024
Company in FDA record
Bio-Rad Laboratories
FDA classification date
Mar 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.

Source facts

Action
On February 7, 2024, Bio-Rad issued their initial "Urgent Medical Device Correction" notification via E-Mail. On February 14, 2023, Bio RAD Laboratories issued a revised "Urgent Medical Device Correction" notification via E-Mail. Bio-Rad asked consignees to take the following actions: 1. Confirm past patient results against correct values for Catalog No. 12005560, Batch Nos. 64562656 and 64571846 when provided with corrected calibrator values by Bio-Rad Laboratories. 2. If all the controls values fall within the corrected ranges, no further action is required. If one or more controls are outside the corrected ranges, that particular run should be considered as failed and the results for the patient samples tested within that run should be considered invalid. 3. Please go to Bio-Rad.com for the updated Certificate of Analysis. 4. Please ensure this notice is passed to all those who need to be aware within your organization or to any organization where the impacted devices have been transferred. 5. Please complete and return the attached response form as soon as possible so that we are assured you have received this important communication. 6. This notice has been reported to the appropriate Regulatory Agencies. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.
Initiated
2024-02-15
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of PA, SD, TX, FL, MN.
Recall number
Z-1358-2024
Recall status
Open, Classified
CDRH record ID
206138
FDA classified
2024-03-22
Recall event number
94020

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗