Skip to content

Public FDA record

Z-1359-2012: OneTouch Verio IQ Blood Glucose Monitoring System Kit; One Touch Verio IQ Blood Glucose IQ Blood Glucose

Recall record Z-1359-2012 for Lifescan Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1359-2012
Company in FDA record
Lifescan Inc
FDA classification date
Apr 2, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An error was discovered in the OneTouch IQ meter software so that the meter turns itself off when a user attempts to view results in the "Results Log" when the log has 256 or a multiple of 256 items to display.

Source facts

Action
Lifescan sent an Urgent Medical Device Voluntary Recall letter dated February 21, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1. Identify and sequester all lots of OneTouch Verio"IQ Meters they have in inventory. UPC No. 353885007702 LifeScan Part No. 022267 NDC No. 53885026701 Note: Only OneTouch Verio"IQ Meters are subject to this notice. Please continue to sell all other OneTouch Brand products including OneTouch Verio" Test Strips and all OneTouch Ultra Brand Meters, Test Strips and Ancillary Products. 2. Communicate this replacement program to their distribution centers and retail stores that have received OneTouch Verio"IQ Meters. Request that they return only OneTouch Verio"IQ Meters per their normal return procedures. 3. Once you received all of the OneTouch Verio"IQ Meters to be returned, call Inmar Corp. at 336-631-7744 for a Returned Goods Authorization (RGA) number and product return instructions. When product has been returned, a credit memo for an amount equivalent to the invoiced price of the returned product. No deductions will be allowed. If you have any questions about this action, please contact your LifeScan Sales Representative or call the LifeScan Customer Experience Line at: 1-800-600-7226, Option 1. Customers were advised that they can continue use of their affected product until an updated meter is provided free of charge. For questions regarding this recall call 408-942-5937.
Initiated
2012-02-21
Root cause
Software design
Terminated
2013-02-05
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Recall number
Z-1359-2012
Recall status
Terminated
CDRH record ID
107697
FDA classified
2012-04-02
Recall event number
61272

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗