Skip to content

Public FDA record

Z-1360-2022: Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO)

Recall record Z-1360-2022 for Isopure Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1360-2022
Company in FDA record
Isopure Corp
FDA record creation date
Jul 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Isopure received a report of a hazardous incident involving a device which had experienced an electrical short that resulted in a thermal event, damaging some of the device components. The firm is alerting customers to electrical safety measures that can be taken to prevent such events from occurring.

Source facts

Action
Firm began to call customers on April 28, 2022, to determine if the IsoMix Pro is still in use at each location. The firm will also be sending an email out to all clinics along with an updated Safety Notice. Once information is gathered, service technicians will visit each clinic to assess the devices. Customers will be provided with a video link as part of an inspection packet on proper use of a torque screwdriver. Customers should follow the provided guidance documents. Customers should always replace any depleted or blown fuses with the correct size. Customers are also instructed to follow the recommended preventive maintennace schedule for the device.
Initiated
2022-04-28
Root cause
Other
Distribution
Domestic distribution to AL AZ CO DE FL GA IN KS KY LA MA MD ME MN MO NC NY OK PA TX UT VA WV No foreign distribution.
Recall number
Z-1360-2022
Recall status
Open, Classified
CDRH record ID
194283
FDA record created
2022-07-11
Recall event number
90389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗