Open original FDA record ↗Recall record
- FDA identifier
- Z-1360-2023
- Company in FDA record
- Coloplast Manufacturing US, LLC
- FDA classification date
- Apr 12, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Source facts
- Action
- The firm issued an URGENT MEDICAL DEVICE VOLUNTARY RECALL notice on 03/01/2023 by UPS next day air. The notice explained the issue and the hazard to health. The notice requests the removal and return of the affected product. Distributors are instructed to notify their customers. For further questions, please contact the Coloplast Customer Service team directly at (800) 258-3476, or by email urology@coloplast.com.
- Initiated
- 2023-03-01
- Root cause
- Process change control
- Distribution
- Worldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.
- Recall number
- Z-1360-2023
- Recall status
- Open, Classified
- CDRH record ID
- 198899
- FDA classified
- 2023-04-12
- Recall event number
- 91800