Skip to content

Public FDA record

Z-1361-2016: Neurotherm NT2000iX RF Generator Model Number: RFG-NT -2000iX

Recall record Z-1361-2016 for Neurotherm, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1361-2016
Company in FDA record
Neurotherm, Inc.
FDA classification date
Apr 9, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.

Source facts

Action
Neurotherm notified consignees via telephone and guided through the process to update the software manually. The calls were initiated by Neurotherm representatives on October 09,2013.
Initiated
2013-10-09
Root cause
Software Manufacturing/Software Deployment
Terminated
2016-04-14
Distribution
CA, CO, IA, ME, NJ, NY, SC, TX, WI
Recall number
Z-1361-2016
Recall status
Terminated
CDRH record ID
144151
FDA classified
2016-04-09
Recall event number
73410

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗