Skip to content

Public FDA record

Z-1361-2026: Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H

Recall record Z-1361-2026 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1361-2026
Company in FDA record
Steris Corporation
FDA classification date
Feb 12, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.

Source facts

Action
All affected Customers have been contacted by a STERIS Service Technician arranging an onsite inspection of their unit(s) to locate and replace the affected MNA(s). Users are reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Diksha Chechi, Quality Manager at 510-439-4500.
Initiated
2025-05-23
Root cause
Process control
Distribution
US Nationwide distribution in the states of FL, NC, NY, OH, TN, TX, and VA.
Recall number
Z-1361-2026
Recall status
Open, Classified
CDRH record ID
218080
FDA classified
2026-02-12
Recall event number
98266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗