Reason stated in the source
The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing.
Source facts
- Action
- Maquet sent a Product Recall Urgent-Medical Device Corrective Action letter dated December 22, 2011 to all affected customers. The letter identified the affected products, patient risk, product issue and actions to be taken. Customers were instructed to complete and return the attached Customer Fax Response form as soon as possible. Maquet Field Service Technicians will update all affected products at all the concerned customer sites. Customers are advised untill Maquet has updated all the affected products (all of which are still under warranty), inform all the concerned biomedical staff about the content of this Device Correction Letter. For technical support call 1-888-627-8383.
- Initiated
- 2011-12-22
- Root cause
- Device Design
- Terminated
- 2013-10-22
- Distribution
- USA Nationwide Distribution
- Recall number
- Z-1362-2012
- Recall status
- Terminated
- CDRH record ID
- 107806
- FDA classified
- 2012-04-03
- Recall event number
- 61327