Z-1362-2016: USE (Ebony) PTA .014" rapid exchange (RX) Peripheral Dilatation Catheter. Distributed by US Endovascular

Recall record Z-1362-2016 for Us Endovascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1362-2016
Company in FDA record
Us Endovascular
FDA record creation date
Apr 11, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product may kink during use rendering the product unusable.

Source facts

Action
NATEC Medical ltd.sent an Urgent Medical Device Advisory Notice dated March 18, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Warnings: This device should be used only by physicians who are experienced and have a thorough understanding of the clinical and technical aspects of angioplasty. What action do you need to take: 1. Always use a guiding catheter when contralateral access is needed. 2. Always insert cautiously the PTA catheter inside the guiding catheter, and if any resistance is felt, the cause should be determined via fluoroscopy. 3. For contralateral procedure, the USE014OTW has to be preferred to Ebony 014 Rx. NATEC Medical Ltd and our distributor US Endovascular are committed to keeping you and your healthcare professional informed of issues and solutions concerning our products and services. Should you have any questions regarding this or any other matter regarding your USE 014Rx PTA catheter, please do not hesitate to contact our distributor: US Endovascular at (314) 341-4930.
Initiated
2016-02-11
Root cause
Device Design
Terminated
2017-11-30
Distribution
US Distribution to the states of : AR, CA, FL, LA, MO, MI, MS, NY, TN, and TX.
Recall number
Z-1362-2016
Recall status
Terminated
CDRH record ID
143315
FDA record created
2016-04-11
Recall event number
73171

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗