Open original FDA record ↗Recall record
- FDA identifier
- Z-1362-2018
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Apr 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
Source facts
- Action
- The customers were contacted initially by the Urgent Medical Device Recall letter on 03/028/2018 delivered by email and US Mail. Beckman Coulter will schedule a sites visit to verify the reporting units of measurement for casts for the laboratory within 90 days of this recall initiation. The use of Casts/HPF is not a common setting worldwide.
- Initiated
- 2018-03-02
- Root cause
- Use error
- Terminated
- 2023-07-06
- Distribution
- Worldwide
- Recall number
- Z-1362-2018
- Recall status
- Terminated
- CDRH record ID
- 162661
- FDA record created
- 2018-04-11
- Recall event number
- 79566