Open original FDA record ↗Recall record
- FDA identifier
- Z-1362-2026
- Company in FDA record
- Wilson-Cook Medical Inc.
- FDA classification date
- Feb 13, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
Source facts
- Action
- On 12/18/2025, the firm priority mailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm is removing Teslatome Bipolar Sphincterotomes from the market because they are nonconforming. Customers are instructed to: 1. Examine inventory immediately to determine if you have affected devices. Immediately cease use of the affected devices and quarantine all affected devices. 2. Return the affected devices to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a credit. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected devices on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. For questions, contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
- Initiated
- 2025-12-18
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Distribution to states of: CA, CO, IN, and NC.
- Recall number
- Z-1362-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217885
- FDA classified
- 2026-02-13
- Recall event number
- 98202