Skip to content

Public FDA record

Z-1362-2026: Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842

Recall record Z-1362-2026 for Wilson-Cook Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1362-2026
Company in FDA record
Wilson-Cook Medical Inc.
FDA classification date
Feb 13, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.

Source facts

Action
On 12/18/2025, the firm priority mailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm is removing Teslatome Bipolar Sphincterotomes from the market because they are nonconforming. Customers are instructed to: 1. Examine inventory immediately to determine if you have affected devices. Immediately cease use of the affected devices and quarantine all affected devices. 2. Return the affected devices to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a credit. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected devices on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. For questions, contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
Initiated
2025-12-18
Root cause
Under Investigation by firm
Distribution
U.S. Distribution to states of: CA, CO, IN, and NC.
Recall number
Z-1362-2026
Recall status
Open, Classified
CDRH record ID
217885
FDA classified
2026-02-13
Recall event number
98202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗