Skip to content

Public FDA record

Z-1363-04: Amplicor [3] AV-HRP. (Avidin-HRP (Horseradish Peroxidase) BGG Conjugate.) Kit: Amplicor IC Detection Kit

Recall record Z-1363-04 for Roche Molecular Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1363-04
Company in FDA record
Roche Molecular Systems Inc
FDA classification date
Aug 19, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.

Source facts

Action
With respect to U.S distribution, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution center, Roche Diagnostics Corporation of Indianapolis, IN. With respect to overseas distributors, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution centers. Roche stated that the local safety officer is responsible to communicate the advisory both to the customers/end-users and the regulatory authority of the country. On July 6, 2004, a Revised Important Product Advice Notice was sent out to affiliates requiring the immediate discontinuation of the use of the affected kits, and destruction of any kits remaining in customers hands.
Initiated
2004-07-07
Root cause
Other
Terminated
2005-01-07
Distribution
♦ Amplicor CT/NG CT Detection Kit, Lot No.E12231, Material No.20759392018, Expiration 2/2005 , Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor CT/NG NG Detection Kit, Lot No.E12647, Material No.20759406018, Expiration 2/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor HCV Detection Kit, v2.0, Lot No.F01023, Material No.21118439018, Expiration 5/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor IC Detection Kit, 96 tests, Lot No.E09554, Material No.20751952018, Expiration 5/2005, Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. Export-U.S Distributed Products: AmpliCap HCV Monitor Test, Lot No.E09193, Material No.21111337018, Expiration 8/2004 , Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan. Amplicor IC Detection Kit, Lot No.E13115, Material No.20763306122, Expiration 7/2005, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada; Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. Amplicor M. avium Detection Kit, Lot No.E12990, Material No.20759880018, Expiration 11/2004, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan. Amplicor MTB Detection Kit, Lot No.E13307, Material No.20757462122, Expiration 9/2004, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. Amplicor NG Detection Kit, Lot No.E10894, Material No.20744026122, Expiration 2/2005, Shipped To: Roche Diagnostics KK, Nippon Roche KK, Kamakura Factory 200, Aza Sotokochi, Kajiwara, Kamakura City, Kanagawa Pref, Japan ; Shipped To: Roche Diagnostics Canada, 201 Boulevard Armand-Frappier, Laval QC H7V 4A2, Canada; Shipped To: Roche Diagnostics GmbH, GX-OSW, Sandhoferstr. 116, D-68305 Mannheim. There are no Govt. accounts.
Recall number
Z-1363-04
Recall status
Terminated
CDRH record ID
34125
FDA classified
2004-08-19
Recall event number
29566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗