Open original FDA record ↗Recall record
- FDA identifier
- Z-1363-2016
- Company in FDA record
- Natus Medical Incorporated
- FDA classification date
- Apr 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.
Source facts
- Action
- Natus of Seattle, WA began sending the Urgent Field Safety Notice NatalCare LX Incubator letter (DOC-017939A) to consignees on 1/8/16. A repair kit to correct this issue is now available for repair of affected units. Consignees should complete the NatalCare LX Incubator Oval Port Hinge Repair Kit Order Form and FAX it 847-949-6031 or scan and email this form to Natus_Quality_Programs@natus.com. Alternately you can telephone Natus Technical Service at 888-496-2887 and order the kit from a specialist. Natus will send the kit(s) free-of-charge to the address you specify. Each upgrade will require up to - 1 hour of time and require simple hand tools. If you have any questions, please contact Technical Service at 888-496-2887 or send email to: customer.service@natus.com
- Initiated
- 2016-01-08
- Root cause
- Process control
- Terminated
- 2017-02-10
- Distribution
- Distribution US nationwide and Pakistan.
- Recall number
- Z-1363-2016
- Recall status
- Terminated
- CDRH record ID
- 142443
- FDA classified
- 2016-04-11
- Recall event number
- 72894