Z-1363-2018: ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003

Recall record Z-1363-2018 for Orthofix, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1363-2018
Company in FDA record
Orthofix, Inc
FDA record creation date
Apr 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Source facts

Action
The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
Initiated
2018-02-23
Root cause
Nonconforming Material/Component
Terminated
2019-06-18
Distribution
US
Recall number
Z-1363-2018
Recall status
Terminated
CDRH record ID
162480
FDA record created
2018-04-11
Recall event number
79498

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗