Skip to content

Public FDA record

Z-1363-2020: ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446

Recall record Z-1363-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1363-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Jan 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.

Source facts

Action
On December 20, 2019, the recalling firm distributed Urgent Medical Device Recall letters to affected customers. Customers were provided with instructions for determining whether the Lipase reagent cartons in their inventory were impacted. The next reagent lot is expected to be available by February 2020. Customers were asked to review the letter with their Medical Director, review their inventory for affected products, and assess their laboratory's replacement needs as applicable. A customer response form was included with the recall communication so that customers may confirm that they have received the notification as well as report the quantity of replacement product required.
Initiated
2019-12-20
Root cause
Under Investigation by firm
Terminated
2020-12-21
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CT, FL, GA, IA, KS, MD, MI, NJ, NM, NY, OH, OK, PA, TN, TX, WA.
Recall number
Z-1363-2020
Recall status
Terminated
CDRH record ID
179303
FDA classified
2020-01-30
Recall event number
84768

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗