Open original FDA record ↗Recall record
- FDA identifier
- Z-1364-2018
- Company in FDA record
- Orthofix, Inc
- FDA record creation date
- Apr 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Source facts
- Action
- The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
- Initiated
- 2018-02-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-06-18
- Distribution
- US
- Recall number
- Z-1364-2018
- Recall status
- Terminated
- CDRH record ID
- 162481
- FDA record created
- 2018-04-11
- Recall event number
- 79498