AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Source facts
Action
Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.
Initiated
2021-08-03
Root cause
Labeling Change Control
Terminated
2024-04-10
Distribution
US Nationwide distribution in the states of NC and WV.
Recall number
Z-1364-2022
Recall status
Terminated
CDRH record ID
194345
FDA record created
2022-07-11
Recall event number
90411