Skip to content

Public FDA record

Z-1365-2013: MEDITECH Bi-Directional interactive Communication Analyzer Interface; Integrates with MEDITECH's Labora

Recall record Z-1365-2013 for Medical Information Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1365-2013
Company in FDA record
Medical Information Technology, Inc.
FDA classification date
May 22, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect transmission of laboratory results.

Source facts

Action
Beginning on May 7, 2013, MEDITECH distributed a notification to MAGIC and Client Server customers who have the MEDITECH Bi-Directional Interactive Communication Analyzer Interface in use with Vitros 3600, 4600, or 5600 Analyzer in TEST and/or LIVE. This notification was made via remailed task updates that can be printed by the customer. Task updates sent via e-mail are immediately transmitted to the customer. This notification informs customers of the potential patient safety issue that could result in reporting inaccurate qualitative results. Customers will receive a corrective change, in which the previous test's results will no longer be associated with the test not performed. Until customers receive the corrective change, they are instructed to monitor all No Result values from their analyzer and confirm that the result in MEDITECH matches. For further questions please call 781-821-3000, ext. 5742.
Initiated
2013-05-07
Root cause
Device Design
Terminated
2014-10-01
Distribution
Worldwide Distribution - USA, including Puerto Rico and the country of Canada.
Recall number
Z-1365-2013
Recall status
Terminated
CDRH record ID
118048
FDA classified
2013-05-22
Recall event number
65158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗