Z-1365-2020: NUCLISENS Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 692

Recall record Z-1365-2020 for BIOMERIEUX - Grenoble. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1365-2020
Company in FDA record
BIOMERIEUX - Grenoble
FDA classification date
Feb 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.

Source facts

Action
Consignees notified via FedEx Express 2 day delivery service with recall letter dated 10 January 2020. The letter identified affected product, stated reason for recall, and requested product to not be used for testing of whole blood samples containing blood and other specimens containing blood. Samples with no hemoglobin are not impacted by the issue and can be used with expected performance. A Recall Acknowledgement Form was requested to be completed and returned via fax.
Initiated
2020-01-10
Root cause
Equipment maintenance
Terminated
2022-04-14
Distribution
No US distribution. No Mexican consignees. International distribution to South Africa, Uruguay, Tanzania, Romania, Netherlands, Macao, Italy, India, Hong Kong, United Kingdom, France, Spain, Germany, China, Chile, Canada, Belgium, Australia, Argentina, Curacao.
Recall number
Z-1365-2020
Recall status
Terminated
CDRH record ID
179301
FDA classified
2020-02-11
Recall event number
84766

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗