Skip to content

Public FDA record

Z-1365-2022: Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Recall record Z-1365-2022 for Draeger Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1365-2022
Company in FDA record
Draeger Medical Systems, Inc.
FDA record creation date
Jul 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.

Source facts

Action
On May 20, 2022, the firm began to send Important Field Safety Notice letters to affected customers. Customers were instructed to avoid using the ventilator portion of the system due to the potential for enriched ambient air rising beyond safe levels until a replacement part can be installed and tested. The firm will send customers a replacement kit and testing which will be completed at no charge to the customer. Draeger Service Representatives will call customers and schedule a service visit to update affected systems.
Initiated
2022-05-20
Root cause
Under Investigation by firm
Distribution
No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
Recall number
Z-1365-2022
Recall status
Open, Classified
CDRH record ID
194359
FDA record created
2022-07-12
Recall event number
90423

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗