Open original FDA record ↗Recall record
- FDA identifier
- Z-1366-2013
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA classification date
- May 22, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
DePuy Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of one lot of the AltrX" liner because the product was incorrectly labeled on the box. The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product enclosed was a 36IDX56OD Neutral Liner. Three ( 3 ) complaints were received in January 2012.
Source facts
- Action
- DePuy Orthopaedics sent an URGENT INFORMATION- MEDICAL DEVICE RECALL NOTICE dated March 7, 2012 via email to all affected distributors and customers. The letter identified the affected product, problem and actions to be taken. Distributors and customers are instructed to cease further distribution or use or of recalled product and to return them to DePuy. The sales representatives are expected to aid customer in returning of recalled product. Distributors and sales representatives are to return the recalled product through the normal DePuy Returns process. Reconciliation form should be completed and returned to your DePuy Sales Representative or faxed to 574-372-7567. For questions about recall information contact the Manager of Customer Quality 574-372-7333.
- Initiated
- 2012-03-07
- Root cause
- Packaging process control
- Terminated
- 2013-05-24
- Distribution
- Worldwide Distribution - USA Nationwide including the state of VA and the country of Ireland.
- Recall number
- Z-1366-2013
- Recall status
- Terminated
- CDRH record ID
- 107791
- FDA classified
- 2013-05-22
- Recall event number
- 61315