Skip to content

Public FDA record

Z-1366-2022: Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Recall record Z-1366-2022 for Wright Medical Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1366-2022
Company in FDA record
Wright Medical Technology, Inc.
FDA record creation date
Jul 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

Source facts

Action
The firm sent an URGENT MEDICAL DEVICE RECALL notice on beginning 06/14/2022 by letter (FedEx). The letter discussed the issue and potential risk, an requested the following actions be taken: "1. Immediately check your internal inventory to locate the product listed on the attached business reply form and remove them from their point of use. 2. Return the enclosed business reply form by email to confirm receipt of this notification/document product segregation. a. Response is required, even if you may not have any physical inventory on site anymore. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations. Including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers. 6. Please inform us of any adverse event and/or report them to the Health/Competent Authorities in accordance with current regulations."
Initiated
2022-06-14
Root cause
Device Design
Distribution
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Recall number
Z-1366-2022
Recall status
Open, Classified
CDRH record ID
194316
FDA record created
2022-07-12
Recall event number
90399

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗