Skip to content

Public FDA record

Z-1367-2008: Gambro cartridge blood tubing set used with Gambro Dasco Phoenix and Cobe Centrysystem 3 (and 3+) Hemodia

Recall record Z-1367-2008 for Gambro Renal Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1367-2008
Company in FDA record
Gambro Renal Products, Inc.
FDA classification date
Mar 25, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Kinked tubing on hemodialysis device may cause hemolysis.

Source facts

Action
On September 10, 2007, the firm sent an Advisory Notice to all consignees alerting them that a kink in the tubing can result in physical hemolysis. The firm included with the Notice a Quick Reference Guide that provides drawing of the proper installation of the Phoenix and describes how to prevent bloodline kinking. A customer reply form was also included. The firm is in the process of making labeling changes that will be released with the updated software revisions.
Initiated
2007-09-10
Root cause
Device Design
Terminated
2012-01-03
Distribution
Nationwide, including 17 VA facilities and 1 military facility. Foreign distribution to Canada, Korea, Mexico.
Recall number
Z-1367-2008
Recall status
Terminated
CDRH record ID
66565
FDA classified
2008-03-25
Recall event number
45950

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗