Kinked tubing on hemodialysis device may cause hemolysis.
Source facts
Action
On September 10, 2007, the firm sent an Advisory Notice to all consignees alerting them that a kink in the tubing can result in physical hemolysis. The firm included with the Notice a Quick Reference Guide that provides drawing of the proper installation of the Phoenix and describes how to prevent bloodline kinking. A customer reply form was also included. The firm is in the process of making labeling changes that will be released with the updated software revisions.
Initiated
2007-09-10
Root cause
Device Design
Terminated
2012-01-03
Distribution
Nationwide, including 17 VA facilities and 1 military facility. Foreign distribution to Canada, Korea, Mexico.
Recall number
Z-1367-2008
Recall status
Terminated
CDRH record ID
66565
FDA classified
2008-03-25
Recall event number
45950