Skip to content

Public FDA record

Z-1367-2013: SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any f

Recall record Z-1367-2013 for Separation Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1367-2013
Company in FDA record
Separation Technology, Inc.
FDA classification date
May 23, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.

Source facts

Action
On March 13, 2013, Separation Technology, Inc. sent their customers letters instructing them to complete and return the attached Acknowledgement & Receipt form via fax to (407) 788-3677 and requesting the return of the SlidePrep Plus Cytology Centrifuges to STI for correction. For any questions regarding this recall, call STI Customer Service at (800) 777-6668.
Initiated
2013-03-13
Root cause
Device Design
Terminated
2013-12-24
Distribution
Worldwide Distribution-USA (nationwide) including the states of PA, NC, IL, KS, and FL, and the countries Canada and Turkey.
Recall number
Z-1367-2013
Recall status
Terminated
CDRH record ID
117450
FDA classified
2013-05-23
Recall event number
64718

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗