Z-1367-2022: Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic

Recall record Z-1367-2022 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1367-2022
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA record creation date
Jul 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

Source facts

Action
Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311
Initiated
2022-05-26
Root cause
Incorrect or no expiration date
Distribution
Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.
Recall number
Z-1367-2022
Recall status
Open, Classified
CDRH record ID
194342
FDA record created
2022-07-12
Recall event number
90408

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗