- Recall classification
- Class II
Reason stated in the source
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.
Source facts
- Action
- Roche Diagnostics issued on 2/04/19 an "URGENT MEDICAL DEVICE CORRECTION"(UMDC) letter to its customers via UPS Ground (receipt required). The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: " When using the automated bottle change function, ensure Good Laboratory Practices (GLP) are being followed (i.e., all QC results for the current and standby bottles are in range and are not flagged prior to testing patient samples). " Complete the attached fax form (TP-00548) and fax this form to 1-866-608-3939 or email it to Roche10042@stericycle.com. " When a new software version is available, a Roche representative will contact you to schedule the software update. " File this UMDC for future reference. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-526-2272 if you have questions
- Initiated
- 2019-02-04
- Root cause
- Software design
- Terminated
- 2020-09-16
- Distribution
- US Distribution to states of: AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN and TX.
- Recall number
- Z-1368-2019
- Recall status
- Terminated
- CDRH record ID
- 171034
- FDA record created
- 2019-05-17
- Recall event number
- 82139