Open original FDA record ↗Recall record
- FDA identifier
- Z-1368-2021
- Company in FDA record
- Medtronic Neuromodulation
- FDA record creation date
- Apr 7, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.
Source facts
- Action
- Beginning 9-Mar-2021, Medtronic representatives began verbally notifications in order to inform impacted physicians of the issues and provide the mitigations outlined in the letters. Beginning 16-Mar-2021, Medtronic notified its consignees via mail, email, fax, and/or personal visit by Medtronic personnel. The notice described the problems and provided a workaround to use until a software update is available.
- Initiated
- 2021-03-09
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide distribution in Puerto Rico and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, Lithuania, Luxembourg, Malaysia, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom.
- Recall number
- Z-1368-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186320
- FDA record created
- 2021-04-07
- Recall event number
- 87569