Z-1368-2024: InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinu

Recall record Z-1368-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1368-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Mar 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be updated to reflect the instructions in the English IFU

Source facts

Action
Olympus issued Urgent Medical Device Corrective Action letter on 1/16/24 to the Distributor-OLYMPUS MEDICAL SYSTEMS CORP LOGITE Tokyo, Japan for distribution to their consignees. Letter states reason for recall, health risk and action to take: Our records indicate that your facility has purchased one or more of the affected InstaClear Sheaths. Olympus requests you to take the following actions: 1. Inspect your inventory and identify any devices with the model names specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Add a copy of the enclosed addendum with your remaining inventory. You may continue to use the products, but this should be done in accordance with the attached labeling addendum. 2. Carefully read the content of this Medical Device Correction Action as well as the attached labelling Addendum . 3. Ensure all current and future personnel, including clinical staff, are completely knowledgeable and thoroughly aware of the contents. The Addendum provides the updated instructions for proper setup of the InstaClear Sheath. 4. Olympus requests that you acknowledge receipt of this letter [local facility method/contact]. 5. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Actions to be taken by the company: Olympus will update the labelling to include the improved instructions regarding proper installation and setup of the InstaClear Sheath. Additionally, your local Olympus contact will reach out to schedule an onsite educational training on setup of the InstaClear device.
Initiated
2024-01-16
Root cause
Under Investigation by firm
Distribution
International distribution to the country of Japan.
Recall number
Z-1368-2024
Recall status
Open, Classified
CDRH record ID
206240
FDA classified
2024-03-22
Recall event number
94067

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗