Z-1368-2026: Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A P

Recall record Z-1368-2026 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1368-2026
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Feb 13, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pumps were released without full testing being performed, including occlusion alarm testing.

Source facts

Action
Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.
Initiated
2026-01-13
Root cause
Employee error
Distribution
States: Ohio Countries: United States
Recall number
Z-1368-2026
Recall status
Open, Classified
CDRH record ID
218214
FDA classified
2026-02-13
Recall event number
98306

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗