Open original FDA record ↗Recall record
- FDA identifier
- Z-1369-2008
- Company in FDA record
- Advanced Sterilization Products
- FDA classification date
- Aug 8, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid opening, and c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).
Source facts
- Action
- Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.
- Initiated
- 2008-02-13
- Root cause
- Other
- Terminated
- 2011-11-29
- Distribution
- Worldwide Distribution --- including USA and countries of Canada and Japan.
- Recall number
- Z-1369-2008
- Recall status
- Terminated
- CDRH record ID
- 68458
- FDA classified
- 2008-08-08
- Recall event number
- 46906