Philips Medical Systems North America Co. Phillips
FDA classification date
Mar 22, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
Source facts
Action
On 11/5/10, Philips began sending the Electronic Product Radiation Warning notification letter, dated November 4, 2010, to their consignees.
The consignees were informed that the middle pedal of the foot switch of the Allura FD10, FD20, CV20 systems, and the Allura Field Extension series does not not always become inactive (sticks) when pedal is released.
The function of the middle pedal on the foot switch is freely programmable by the customer. If it is programmed for exposure and if the pedal does not release by itself, the patient will be exposed to X-ray longer than intended.
The consignees were advised to check if their footswitch is possibly affected. Footswitch identification 452270000141 is noted. If the pedal does not release by itself press the red emergency stop button on the system.
Consignees can call Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 72200189.
Initiated
2010-11-04
Root cause
Pending
Terminated
2011-07-21
Distribution
Worldwide Distribution -- US, including states of CA, IL, MA, NC, and OC.
Recall number
Z-1369-2011
Recall status
Terminated
CDRH record ID
97839
FDA classified
2011-03-22
Recall event number
57106