Open original FDA record ↗Recall record
- FDA identifier
- Z-1369-2019
- Company in FDA record
- Cook Inc.
- FDA record creation date
- May 17, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
Source facts
- Action
- Cook issued MEDICAL DEVICE CORRECTION letters via courier to US Consignees on April 22, 2019. The letter provides reason for recall, products affected, health risk, action to take: Understand that stent knotting is a potential complication associated with use of the Kwart Retro-Inject" Stent Set and Salle Intraoperative Pyeloplasty Stent Set and should be considered if significant resistance is encountered during attempts at removal. 2. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 3. If you have affected product in inventory, you may continue to use these products. Cook Medical is not requesting product be returned. 4. Complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter.5. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred. Questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
- Initiated
- 2019-04-22
- Root cause
- Error in labeling
- Terminated
- 2020-05-05
- Distribution
- Nationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jordan Slovakia South Africa United Arab Emirates Slovenia Lithuania Morocco Romania Qatar Bulgaria Libya Saudi Arabia Turkey Ukraine Thailand
- Recall number
- Z-1369-2019
- Recall status
- Terminated
- CDRH record ID
- 172723
- FDA record created
- 2019-05-17
- Recall event number
- 82795