Z-1369-2019: Cook Kwart Retro-Inject Ureteral Stent Set: PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr 003500 G14916; 6.0 Fr

Recall record Z-1369-2019 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1369-2019
Company in FDA record
Cook Inc.
FDA record creation date
May 17, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.

Source facts

Action
Cook issued MEDICAL DEVICE CORRECTION letters via courier to US Consignees on April 22, 2019. The letter provides reason for recall, products affected, health risk, action to take: Understand that stent knotting is a potential complication associated with use of the Kwart Retro-Inject" Stent Set and Salle Intraoperative Pyeloplasty Stent Set and should be considered if significant resistance is encountered during attempts at removal. 2. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 3. If you have affected product in inventory, you may continue to use these products. Cook Medical is not requesting product be returned. 4. Complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter.5. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred. Questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
Initiated
2019-04-22
Root cause
Error in labeling
Terminated
2020-05-05
Distribution
Nationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jordan Slovakia South Africa United Arab Emirates Slovenia Lithuania Morocco Romania Qatar Bulgaria Libya Saudi Arabia Turkey Ukraine Thailand
Recall number
Z-1369-2019
Recall status
Terminated
CDRH record ID
172723
FDA record created
2019-05-17
Recall event number
82795

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗