Z-1369-2021: Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary

Recall record Z-1369-2021 for Hologic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1369-2021
Company in FDA record
Hologic, Inc
FDA record creation date
Apr 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient results.

Source facts

Action
On February 26, 2021, Hologic, issued "Urgent: Medical Device Recall Notification" to affected consignees via FedEx. In addition, to informing consignees about the recall, they asked customers to take the following actions: Laboratory Action: A. Please immediately discontinue use of the Panther Fusion Tube Tray lots listed in Table 1. B. Please immediately check your inventory and segregate the Panther Fusion Tube Tray lots listed in Table 1. C. The number of tube tray boxes shipped to your site is identified on the accompanying document. (GBS customers will receive a blank form and will need to fill in lot number and quantity.) - Please fill out the documents completely, including the number of boxes of the specified lots remaining in inventory (whether unopened or partial boxes). - After completing the document, destroy the segregated inventory. - Sign the document to signify compliance and return it as instructed on the document. - Please do this even if you do not have any of the affected trays remaining in your inventory. D. Hologic will ship replacement product to your site based on the number of boxes of the specified lots you list as remaining in your inventory on the document you return. E. Please ensure that appropriate laboratory personnel are notified of the information herein. F. If you have any questions or concerns about this notification or about the replacement product, please contact Hologic Technical Support at 1888-484-474- or molecularsupport@hologic.com
Initiated
2021-02-26
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CT, DC, FL, GA, HI, IL, IN, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NY, OH, OR, PA, TX, UT, VA, VT, WA, WI, WV and the countries of Austria (AT), Belgium (BE), France (FR), Germany (DE), Great Britain (GB), Italy (IT) Netherlands (NL), Spain (ES), Switzerland (CH),, Canada (CA), Hong Kong (HK), Australia (AU), New Zealand (NZ).
Recall number
Z-1369-2021
Recall status
Open, Classified
CDRH record ID
186269
FDA record created
2021-04-07
Recall event number
87557

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗