Z-1369-2024: Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay o

Recall record Z-1369-2024 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1369-2024
Company in FDA record
Randox Laboratories Ltd.
FDA classification date
Mar 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods

Source facts

Action
Randox Laboratories Ltd, UK ( manufacturer) notified the US Distributor Randox Laboratories USA . The distributor initial contact and the follow up contact via email on 2/20/24. Letter states reason for recall, health risk and action to take: We have realigned the target values in the above lot numbers to the IFCC Reference Standard. Please refer to the table below for the updated calibrator targets. You may experience a positive shift in Quality Control and patient sample recovery. If you are using CVS 2699 batch 600957 lots 1378CY-1382CY please provide proof of scrappage and a replacement calibrator lot will be provided. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com. If further information is required, please contact technical.services@randox.com . In line with the realignment to the Cystatin C Calibrator Series, we are updating the target and range for the Cystatin C controls, CYS5019 and CYS5020. The updated target and range can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
Initiated
2024-02-14
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA, IN.
Recall number
Z-1369-2024
Recall status
Open, Classified
CDRH record ID
206296
FDA classified
2024-03-22
Recall event number
94088

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗