The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
Source facts
Action
Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
Initiated
2005-06-07
Root cause
Other
Terminated
2007-07-02
Distribution
Nationwide.
Recall number
Z-1370-05
Recall status
Terminated
CDRH record ID
39124
FDA classified
2005-08-13
Recall event number
32100