Open original FDA record ↗Recall record
- FDA identifier
- Z-1370-2015
- Company in FDA record
- BD Biosciences, Systems & Reagents
- FDA classification date
- Jun 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Source facts
- Action
- BD Biosciences sent an Important Product Recall letter on March 6, 2015 by certified mail and Field Change Notification (FCN FCB15-01-660312 directing Field Service Engineers to correct the circuit breaker is in process. Customers with questions were instruct to contact BD Customer Support at 1-877-232-8995 (prompt 2) in the United States. For customers outside the US were instructed to contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.
- Initiated
- 2015-03-06
- Root cause
- Mixed-up of materials/components
- Terminated
- 2016-04-04
- Distribution
- Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
- Recall number
- Z-1370-2015
- Recall status
- Terminated
- CDRH record ID
- 134477
- FDA classified
- 2015-06-26
- Recall event number
- 70678