Z-1370-2021: MicroScan WalkAway-40 plus Instrument

Recall record Z-1370-2021 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1370-2021
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Apr 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Source facts

Action
On 09/04/2019, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to its customers via electronic and/or first class postal mail informing them that an investigation revealed that some instruments were manufactured with shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure. Customers were also informed that when a reagent is not dispensed, there is the potential for any of the following to occur: -high probability misidentifications and reporting of erroneous results -panel reads are flagged with exception requiring investigation (e.g. repeat the panel) causing delayed results. -QC failure requiring investigation (e.g. repeat the panel) causing delayed results The reagent draw straw is shorter than the probe level sensor, thus customers will not be notified by LEDs that the reagent system needs attention. Customers are instructed to: 1) Perform daily maintenance and manually verify the reagent volume in the 30mL bottles. Make sure each 30mL reagent bottle has at least 1 1/4 inch (approximately 1/2 full) of reagent or replace or refill bottles if necessary. 2) If the customer experienced an increase of reagents not dispensing, misidentifications and/or QC failures with their instrument, then a retrospective review of patient results is at the discretion of the Laboratory Director. The Recalling Firm will contact customers to schedule an onsite visit. During the visit the Recalling Firm's representative will install the reagent bottle assembly with the correct reagent draw straw length. Customers are asked to share this information with their laboratory staff and retain the recall notification as part of their laboratory Quality System documentation. If the affected systems have been forwarded to another laboratory, please provide a copy of the recall notification to them. Customers are asked to respond within 10 days that they have received
Initiated
2019-09-04
Root cause
Process control
Terminated
2023-01-11
Distribution
U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.
Recall number
Z-1370-2021
Recall status
Terminated
CDRH record ID
186150
FDA record created
2021-04-07
Recall event number
87486

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗