Z-1370-2024: SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs,

Recall record Z-1370-2024 for Alcon Research, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1370-2024
Company in FDA record
Alcon Research, LLC
FDA classification date
Mar 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.

Source facts

Action
On 2/21/24, correction notices were mailed to customers who were asked to do the following: 1) Review your inventory to determine if you have any affected product within your facility. 2) If you have any affected Custom Pak lots in inventory, please use the included stickers to label any affected product. If you have previously reshelved either of the affected Medline Bulb Irrigation Syringe lots, dispose them. 3) All components inside the Pak are intended to be used only for one specific surgical case and should be discarded at the end of the surgical case. 4) Forward the correction notice to all departments within your organization who may be in possession of affected product; and any other organization to which this product may have been transferred. 5) Complete and return the response form via email to Market.Actions@Alcon.com Customers with questions or concerns can call the firm at 1-800-862-5266.
Initiated
2024-02-21
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of OH, WV.
Recall number
Z-1370-2024
Recall status
Open, Classified
CDRH record ID
206318
FDA classified
2024-03-22
Recall event number
94094

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗