Skip to content

Public FDA record

Z-1371-2018: Hamilton-C2 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optional

Recall record Z-1371-2018 for Hamilton Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1371-2018
Company in FDA record
Hamilton Medical AG
FDA record creation date
Apr 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.

Source facts

Action
Customers were notified via letter on about 03/12/2018. The letter identified the affected device and described the issue. Instructions for medical facilities are to inspect the batteries for the battery manufacturing date and replace any batteries that are found to be in use for two years or more. Instructions for determining the battery manufacturing date are included in the Medical Device Field Safety Corrective Action Update attached to the letter. Questions or concerns can be directed to Bret Everett at 817-909-0308 or Hamilton Technical Support at 1-800-426-6331.
Initiated
2018-03-12
Root cause
Other
Distribution
Worldwide Distribution -- USA, including the states of AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MI, MN, MO, MT, NC, ND, NE, NV, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Algeria, Angola, Argentina, Armenia, Australia, Austria, Austria, Azerbaijan, Bangladesh, Belarus, Belgian, Bolivia, Brazil, Bulgaria, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czechia, Denmark, Dominican. Rep., Ecuador, Egypt, El Salvador, Estland, Finland, France, Gabon, Germany, Greece, Guatemala, Holland, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mauritius, Macedonia, Mexico, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Rwanda, Saudi Arabia, Scotland, Serbia, Seychelles, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, Sudan, Suriname, Sweden, Sweden, Switzerland, Syria, Taiwan, Tajikistan, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, UK , Ukraine, Uruguay, Venezuela, Vietnam, and Yemen.
Recall number
Z-1371-2018
Recall status
Open, Classified
CDRH record ID
159861
FDA record created
2018-04-12
Recall event number
78599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.